Have a personal or library account? Click to login
The Calibration Process and Metrological Analysis of a Transducer Used to Measure Two Physical Quantities Cover

The Calibration Process and Metrological Analysis of a Transducer Used to Measure Two Physical Quantities

Open Access
|Jun 2017

Abstract

This article presents a way of calibration of an unconventional two-current circuit, named 2J+2R, which consists of two current sources and two referential resistors connected to the circuit mass. This bridge was used to measure the beam deflection and the temperature increase simultaneously with the use of a pair of metal strain gauges. This paper contains theoretical and corrected (after calibration) processing characteristics of the measurement circuit. Calibration coefficients of both inputs, responsible for measurement of the measured values in the places where the strain gauges are attached, were calculated. Moreover, the standard combined and expanded uncertainties of both calibration coefficients were calculated and an uncertainty budget was made.

DOI: https://doi.org/10.1515/ama-2017-0021 | Journal eISSN: 2300-5319 | Journal ISSN: 1898-4088
Language: English
Page range: 135 - 142
Submitted on: Oct 10, 2016
Accepted on: May 19, 2017
Published on: Jun 15, 2017
Published by: Bialystok University of Technology
In partnership with: Paradigm Publishing Services
Publication frequency: 4 issues per year

© 2017 Adam Idźkowski, Jerzy Gołębiowski, Wojciech Walendziuk, published by Bialystok University of Technology