Table 1

Requirements and recommendations for remote telerobotic ultrasound. FDA: Food and Drug administration; HIPAA: Health Insurance Portability and Accountability Act; DOF: degrees of freedom; MA: medical assistant

REQUIREMENTS AND RECOMMENDATIONS
SonographerState licensing required in New Hampshire, New Mexico, North Dakota, and Oregon
Possible credentialing at operator site, patient site, and the company that develops the technology
NetworkCybersecurity in accordance with FDA guidance documents
HIPAA-compliant encrypted data transmission
Low latency (</ = 250 ms) with minimal packet loss
Video resolution of 1080P at 30 frames per second
Robotic commands over low bandwidth networks (~512 Kbps)
Video, imaging, and audio over high bandwidth (~5 Mbps)
RoboticIdeally 6 DOF motion to duplicate all sonographer hand motions
Regulated force and speed
Sensors in X, Y, and Z axis
Patient SiteMedical assistant (MA) present
Effective two-way communication with sonographer
Patient, MA, and sonographer capacity to stop and withdraw the robotic arm
Sonographer SiteVisual and haptic feedback
Ideally 6 DOF robotic control, mirroring all manual sonographer movements
Control of ultrasound settings and functions
DOI: https://doi.org/10.14797/mdcvj.1645 | Journal eISSN: 1947-6108
Language: English
Page range: 60 - 70
Submitted on: May 26, 2025
Accepted on: Aug 5, 2025
Published on: Oct 7, 2025
Published by: Houston Methodist DeBakey Heart & Vascular Center
In partnership with: Paradigm Publishing Services

© 2025 Wayne L. Monsky, Ryan C. James, Achyuthan Unni Krishnan, Stephen P. Seslar, published by Houston Methodist DeBakey Heart & Vascular Center
This work is licensed under the Creative Commons Attribution-NonCommercial 4.0 License.